Range uncertainty and dose uniformity in proton therapy in the presence of inhomogeneities in tissue and phantom materials / SUMAIRA NAZIR

Swansea University Author: SUMAIRA NAZIR

  • Nazir_Sumaira_PhD_Thesis_Final_Redacted.pdf

    PDF | Redacted version - open access

    Copyright: The author, Sumaira Nazir, 2022.

    Download (7.94MB)

DOI (Published version): 10.23889/SUthesis.59235

Abstract

Proton therapy is an emerging modality for providing radiation treatment to cancer pa-tients. The principal advantages of proton therapy are the reduced total dose deposited into the patient as compared to conventional photon therapy and the finite range of the proton beam. It is considered as a mor...

Full description

Published: Swansea 2022
Institution: Swansea University
Degree level: Doctoral
Degree name: Ph.D
Supervisor: Hugtenburg, Richard
URI: https://cronfa.swan.ac.uk/Record/cronfa59235
Tags: Add Tag
No Tags, Be the first to tag this record!
first_indexed 2022-01-25T09:33:03Z
last_indexed 2022-01-26T04:29:39Z
id cronfa59235
recordtype RisThesis
fullrecord <?xml version="1.0"?><rfc1807><datestamp>2022-01-25T12:08:36.6061364</datestamp><bib-version>v2</bib-version><id>59235</id><entry>2022-01-25</entry><title>Range uncertainty and dose uniformity in proton therapy in the presence of inhomogeneities in tissue and phantom materials</title><swanseaauthors><author><sid>c715c687afda184b3cf6c71a8af28736</sid><firstname>SUMAIRA</firstname><surname>NAZIR</surname><name>SUMAIRA NAZIR</name><active>true</active><ethesisStudent>false</ethesisStudent></author></swanseaauthors><date>2022-01-25</date><abstract>Proton therapy is an emerging modality for providing radiation treatment to cancer pa-tients. The principal advantages of proton therapy are the reduced total dose deposited into the patient as compared to conventional photon therapy and the finite range of the proton beam. It is considered as a more favorable option for optimum treatment outcomes in terms of maximising tumor control probability and minimising normal tissue complications. The depth dose distribution of the proton beam adds an additional degree of freedom to treat-ment planning. The range in tissue is associated with substantial uncertainties triggered by imaging, patient set-up, beam delivery and dose calculations. Therefore, reduction in uncer-tainties would allow to minimise the treatment volume and thus allow a better usage of the protons. However, the presence of sub-millimeter sized heterogeneities, such is pronounced in trabecular bone and lung parenchyma, in the path of the proton beam can cause the Bragg peak degradation with a widening to the distal fall-o&#x2D9;. Additionally, the restricted resolution of a classic CT scan used in treatment planning cannot fully resolve such fine structures, potentially leading to inaccuracy in determination of the range.This work aims to investigate the presence of range uncertainties in proton therapy beams when they penetrate through the sub-millimeter sized heterogeneities. The e&#x2D9;ect of Bragg peak degradation has been demonstrated in bone models with the FLUKA Monte Carlo code and experimental measurements with a 36 MeV proton beam. The bone-substitute material, SAWBONES&#xAE;, ranging in density from 0.088 to 0.48 g/cc, was used to simulate bone heterogeneities. Micro-CT images were obtained of the SAWBONES&#xAE; material and used to construct Monte Carlo models of realistic proton radiotherapy treatments and to benchmark experimental studies. Broadening of the Bragg peak and shifts in the range, as defined by the d20% depth-dose parameter were observed both experimentally and in Monte Carlo models, indicating that such e&#x2D9;ects are in principle, clinically relevant in certain circumstances.Furthermore, a FLUKA Monte Carlo model is benchmarked against the Eclipse treatment planning system (TPS) golden data for proton beam therapy. This project is designed to obtain the proton dose distributions from TPS for a 10 &#xD7;10 &#xD7; 10 cm3 water-filled box. A Monte Carlo analytical model is developed by utilising the information from the TPS to recalculate the dose distribution which are then compared to find any di&#x2D9;erences (if present) for di&#x2D9;erent phantom materials. Due to the lack of any experimental information to measure the normalized depth dose as a function of energy, considering the general behaviour of a monitor chamber, it has been assumed that the treatment planning system has a built-in relationship between the monitor units (MU) and dose delivered. A mathematical formula is developed to find the relationship between monitor units (MU) and dose (E). MU &#x3B1; 1/Ea.The value of a is varied from 0 to 1. It has been observed that the beam non-uniformity calculated by using the relationship " MU = E&#x2212;0.5 " is only 0.15 % for water and 0.43 % for graphite. This non-uniformity in graphite is not severe and it is actually clinically acceptable. This modeling has suggested that the planned dose distributions for water can also be replaced by graphite to a reasonably acceptable standard.</abstract><type>E-Thesis</type><journal/><volume/><journalNumber/><paginationStart/><paginationEnd/><publisher/><placeOfPublication>Swansea</placeOfPublication><isbnPrint/><isbnElectronic/><issnPrint/><issnElectronic/><keywords>range uncertainties, proton therapy</keywords><publishedDay>25</publishedDay><publishedMonth>1</publishedMonth><publishedYear>2022</publishedYear><publishedDate>2022-01-25</publishedDate><doi>10.23889/SUthesis.59235</doi><url/><notes>A selection of third party content is redacted or is partially redacted from this thesis due to copyright restrictions.</notes><college>COLLEGE NANME</college><CollegeCode>COLLEGE CODE</CollegeCode><institution>Swansea University</institution><supervisor>Hugtenburg, Richard</supervisor><degreelevel>Doctoral</degreelevel><degreename>Ph.D</degreename><apcterm/><lastEdited>2022-01-25T12:08:36.6061364</lastEdited><Created>2022-01-25T09:29:02.6560741</Created><path><level id="1">Faculty of Medicine, Health and Life Sciences</level><level id="2">Swansea University Medical School - Medicine</level></path><authors><author><firstname>SUMAIRA</firstname><surname>NAZIR</surname><order>1</order></author></authors><documents><document><filename>59235__22217__89a10395893445199a81f92f758f2b05.pdf</filename><originalFilename>Nazir_Sumaira_PhD_Thesis_Final_Redacted.pdf</originalFilename><uploaded>2022-01-25T12:06:31.3179331</uploaded><type>Output</type><contentLength>8329472</contentLength><contentType>application/pdf</contentType><version>Redacted version - open access</version><cronfaStatus>true</cronfaStatus><documentNotes>Copyright: The author, Sumaira Nazir, 2022.</documentNotes><copyrightCorrect>true</copyrightCorrect><language>eng</language></document></documents><OutputDurs/></rfc1807>
spelling 2022-01-25T12:08:36.6061364 v2 59235 2022-01-25 Range uncertainty and dose uniformity in proton therapy in the presence of inhomogeneities in tissue and phantom materials c715c687afda184b3cf6c71a8af28736 SUMAIRA NAZIR SUMAIRA NAZIR true false 2022-01-25 Proton therapy is an emerging modality for providing radiation treatment to cancer pa-tients. The principal advantages of proton therapy are the reduced total dose deposited into the patient as compared to conventional photon therapy and the finite range of the proton beam. It is considered as a more favorable option for optimum treatment outcomes in terms of maximising tumor control probability and minimising normal tissue complications. The depth dose distribution of the proton beam adds an additional degree of freedom to treat-ment planning. The range in tissue is associated with substantial uncertainties triggered by imaging, patient set-up, beam delivery and dose calculations. Therefore, reduction in uncer-tainties would allow to minimise the treatment volume and thus allow a better usage of the protons. However, the presence of sub-millimeter sized heterogeneities, such is pronounced in trabecular bone and lung parenchyma, in the path of the proton beam can cause the Bragg peak degradation with a widening to the distal fall-o˙. Additionally, the restricted resolution of a classic CT scan used in treatment planning cannot fully resolve such fine structures, potentially leading to inaccuracy in determination of the range.This work aims to investigate the presence of range uncertainties in proton therapy beams when they penetrate through the sub-millimeter sized heterogeneities. The e˙ect of Bragg peak degradation has been demonstrated in bone models with the FLUKA Monte Carlo code and experimental measurements with a 36 MeV proton beam. The bone-substitute material, SAWBONES®, ranging in density from 0.088 to 0.48 g/cc, was used to simulate bone heterogeneities. Micro-CT images were obtained of the SAWBONES® material and used to construct Monte Carlo models of realistic proton radiotherapy treatments and to benchmark experimental studies. Broadening of the Bragg peak and shifts in the range, as defined by the d20% depth-dose parameter were observed both experimentally and in Monte Carlo models, indicating that such e˙ects are in principle, clinically relevant in certain circumstances.Furthermore, a FLUKA Monte Carlo model is benchmarked against the Eclipse treatment planning system (TPS) golden data for proton beam therapy. This project is designed to obtain the proton dose distributions from TPS for a 10 ×10 × 10 cm3 water-filled box. A Monte Carlo analytical model is developed by utilising the information from the TPS to recalculate the dose distribution which are then compared to find any di˙erences (if present) for di˙erent phantom materials. Due to the lack of any experimental information to measure the normalized depth dose as a function of energy, considering the general behaviour of a monitor chamber, it has been assumed that the treatment planning system has a built-in relationship between the monitor units (MU) and dose delivered. A mathematical formula is developed to find the relationship between monitor units (MU) and dose (E). MU α 1/Ea.The value of a is varied from 0 to 1. It has been observed that the beam non-uniformity calculated by using the relationship " MU = E−0.5 " is only 0.15 % for water and 0.43 % for graphite. This non-uniformity in graphite is not severe and it is actually clinically acceptable. This modeling has suggested that the planned dose distributions for water can also be replaced by graphite to a reasonably acceptable standard. E-Thesis Swansea range uncertainties, proton therapy 25 1 2022 2022-01-25 10.23889/SUthesis.59235 A selection of third party content is redacted or is partially redacted from this thesis due to copyright restrictions. COLLEGE NANME COLLEGE CODE Swansea University Hugtenburg, Richard Doctoral Ph.D 2022-01-25T12:08:36.6061364 2022-01-25T09:29:02.6560741 Faculty of Medicine, Health and Life Sciences Swansea University Medical School - Medicine SUMAIRA NAZIR 1 59235__22217__89a10395893445199a81f92f758f2b05.pdf Nazir_Sumaira_PhD_Thesis_Final_Redacted.pdf 2022-01-25T12:06:31.3179331 Output 8329472 application/pdf Redacted version - open access true Copyright: The author, Sumaira Nazir, 2022. true eng
title Range uncertainty and dose uniformity in proton therapy in the presence of inhomogeneities in tissue and phantom materials
spellingShingle Range uncertainty and dose uniformity in proton therapy in the presence of inhomogeneities in tissue and phantom materials
SUMAIRA NAZIR
title_short Range uncertainty and dose uniformity in proton therapy in the presence of inhomogeneities in tissue and phantom materials
title_full Range uncertainty and dose uniformity in proton therapy in the presence of inhomogeneities in tissue and phantom materials
title_fullStr Range uncertainty and dose uniformity in proton therapy in the presence of inhomogeneities in tissue and phantom materials
title_full_unstemmed Range uncertainty and dose uniformity in proton therapy in the presence of inhomogeneities in tissue and phantom materials
title_sort Range uncertainty and dose uniformity in proton therapy in the presence of inhomogeneities in tissue and phantom materials
author_id_str_mv c715c687afda184b3cf6c71a8af28736
author_id_fullname_str_mv c715c687afda184b3cf6c71a8af28736_***_SUMAIRA NAZIR
author SUMAIRA NAZIR
author2 SUMAIRA NAZIR
format E-Thesis
publishDate 2022
institution Swansea University
doi_str_mv 10.23889/SUthesis.59235
college_str Faculty of Medicine, Health and Life Sciences
hierarchytype
hierarchy_top_id facultyofmedicinehealthandlifesciences
hierarchy_top_title Faculty of Medicine, Health and Life Sciences
hierarchy_parent_id facultyofmedicinehealthandlifesciences
hierarchy_parent_title Faculty of Medicine, Health and Life Sciences
department_str Swansea University Medical School - Medicine{{{_:::_}}}Faculty of Medicine, Health and Life Sciences{{{_:::_}}}Swansea University Medical School - Medicine
document_store_str 1
active_str 0
description Proton therapy is an emerging modality for providing radiation treatment to cancer pa-tients. The principal advantages of proton therapy are the reduced total dose deposited into the patient as compared to conventional photon therapy and the finite range of the proton beam. It is considered as a more favorable option for optimum treatment outcomes in terms of maximising tumor control probability and minimising normal tissue complications. The depth dose distribution of the proton beam adds an additional degree of freedom to treat-ment planning. The range in tissue is associated with substantial uncertainties triggered by imaging, patient set-up, beam delivery and dose calculations. Therefore, reduction in uncer-tainties would allow to minimise the treatment volume and thus allow a better usage of the protons. However, the presence of sub-millimeter sized heterogeneities, such is pronounced in trabecular bone and lung parenchyma, in the path of the proton beam can cause the Bragg peak degradation with a widening to the distal fall-o˙. Additionally, the restricted resolution of a classic CT scan used in treatment planning cannot fully resolve such fine structures, potentially leading to inaccuracy in determination of the range.This work aims to investigate the presence of range uncertainties in proton therapy beams when they penetrate through the sub-millimeter sized heterogeneities. The e˙ect of Bragg peak degradation has been demonstrated in bone models with the FLUKA Monte Carlo code and experimental measurements with a 36 MeV proton beam. The bone-substitute material, SAWBONES®, ranging in density from 0.088 to 0.48 g/cc, was used to simulate bone heterogeneities. Micro-CT images were obtained of the SAWBONES® material and used to construct Monte Carlo models of realistic proton radiotherapy treatments and to benchmark experimental studies. Broadening of the Bragg peak and shifts in the range, as defined by the d20% depth-dose parameter were observed both experimentally and in Monte Carlo models, indicating that such e˙ects are in principle, clinically relevant in certain circumstances.Furthermore, a FLUKA Monte Carlo model is benchmarked against the Eclipse treatment planning system (TPS) golden data for proton beam therapy. This project is designed to obtain the proton dose distributions from TPS for a 10 ×10 × 10 cm3 water-filled box. A Monte Carlo analytical model is developed by utilising the information from the TPS to recalculate the dose distribution which are then compared to find any di˙erences (if present) for di˙erent phantom materials. Due to the lack of any experimental information to measure the normalized depth dose as a function of energy, considering the general behaviour of a monitor chamber, it has been assumed that the treatment planning system has a built-in relationship between the monitor units (MU) and dose delivered. A mathematical formula is developed to find the relationship between monitor units (MU) and dose (E). MU α 1/Ea.The value of a is varied from 0 to 1. It has been observed that the beam non-uniformity calculated by using the relationship " MU = E−0.5 " is only 0.15 % for water and 0.43 % for graphite. This non-uniformity in graphite is not severe and it is actually clinically acceptable. This modeling has suggested that the planned dose distributions for water can also be replaced by graphite to a reasonably acceptable standard.
published_date 2022-01-25T04:16:23Z
_version_ 1763754097956093952
score 11.02167