Journal article 412 views
In vitro hemolytic performance of the Realheart® V11C TAH prototype with porcine blood
Ina Laura Perkins ,
Eva Maria Mansisidor ,
Shaikh Faisal Zaman ,
Kyle Mudge ,
Owen Bodger ,
Azad Najar
Artificial Organs, Volume: 47, Issue: 7, Pages: 1208 - 1213
Swansea University Author: Owen Bodger
Full text not available from this repository: check for access using links below.
DOI (Published version): 10.1111/aor.14533
Abstract
Background: Hemolysis testing of new devices to treat heart failure is a regulatory requirement. The ASTM F1841–97 standard for hemolysis testing was developed for continuous flow pumps and does not specify test rig design. When research groups use different methodologies, results are difficult to c...
Published in: | Artificial Organs |
---|---|
ISSN: | 0160-564X 1525-1594 |
Published: |
Wiley
2023
|
Online Access: |
Check full text
|
URI: | https://cronfa.swan.ac.uk/Record/cronfa63882 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Abstract: |
Background: Hemolysis testing of new devices to treat heart failure is a regulatory requirement. The ASTM F1841–97 standard for hemolysis testing was developed for continuous flow pumps and does not specify test rig design. When research groups use different methodologies, results are difficult to compare. Pulsatile flow pump rigs require compliance chambers, and thus, the Aachen rig (Gräf et al) was developed for the pulsatile Reinheart TAH. The study objective was to use this rig to test the early Realheart TAH prototype V11C hemolysis performance compared to literature. Methods: The experimental control was the continuous flow pump BPX-80 (Medtronic) and pooled heparinized porcine blood was used. Results: The mgNIH of BPX-80 and V11C was 5.42 ± 1.47 and 25.20 ± 5.46 mg/100 L, respectively. The NIH ratio of V11C over BPX-80 was 5.5. Conclusion: The absolute and the relative hemolysis of the V11C are lower compared to both the large and small Reinheart TAH devices published values. Pulsatile pumps create more hemolysis in the Aachen rig, and it is not known if this is because how the rig handles pulsatile flow or due to the devices. Future studies will, therefore, use a pulsatile pump such as the SynCardia as clinical comparator and human blood to test the performance of future Realheart TAH prototypes. |
---|---|
Keywords: |
BPX-80; Realheart; hemolysis; porcine; total artificial heart. |
College: |
Swansea University Medical School |
Funders: |
Scandinavian Real Heart AB. |