No Cover Image

Journal article 71 views 15 downloads

Comparability Evaluation of Three Benchtop Glucose Analyzers With the Recently Withdrawn YSI 2300 Stat Plus

Gareth Dunseath Orcid Logo, IULIUS-DUMITRU VATAVU, Steve Luzio Orcid Logo

Journal of Diabetes Science and Technology

Swansea University Authors: Gareth Dunseath Orcid Logo, IULIUS-DUMITRU VATAVU, Steve Luzio Orcid Logo

  • 66252.VoR.pdf

    PDF | Version of Record

    © 2024 Diabetes Technology Society. This article is distributed under the terms of the Creative Commons Attribution 4.0 License.

    Download (249.91KB)

Abstract

Background:We compared the performance of three currently available laboratory benchtop glucose analyzers with the outgoing YSI 2300 Stat Plus.Methods:Plasma samples (100), across a wide glucose concentration range were analysed on the YSI 2500, Randox daytona+ (glucose oxidase) and EKF Biosen in a...

Full description

Published in: Journal of Diabetes Science and Technology
ISSN: 1932-2968 1932-2968
Published: SAGE Publications 2024
Online Access: Check full text

URI: https://cronfa.swan.ac.uk/Record/cronfa66252
Tags: Add Tag
No Tags, Be the first to tag this record!
Abstract: Background:We compared the performance of three currently available laboratory benchtop glucose analyzers with the outgoing YSI 2300 Stat Plus.Methods:Plasma samples (100), across a wide glucose concentration range were analysed on the YSI 2500, Randox daytona+ (glucose oxidase) and EKF Biosen in a single laboratory and compared to the YSI 2300 Stat Plus.Results:All three analyzers showed good agreement with the YSI 2300 Stat Plus, and only a small bias (≤1% YSI 2500 and Randox daytona+, 4.6% EKF Biosen) was observed for each analyzer. None of the three comparator analyzers were affected by either proportional or constant bias, thus no significant differences between the YSI 2300 Stat Plus and the comparator methods were identified.Conclusions:The results from this study suggest all could be considered as suitable reference laboratory glucose analyzers and replacements for the recently withdrawn YSI 2300 Stat Plus.
Keywords: analyzer; glucose; comparison
College: Faculty of Medicine, Health and Life Sciences
Funders: Swansea University